CHAPTER SEVEN
7
Routine cleaning
Complete flushing
Weekly cleaning
Power OFF
For high accuracy and reliability of the device, the following cleaning
procedures should be performed on due times.
7.1. Routine cleaning
When the interval between 2 consecutive measurement analyses is
more than 30 minutes or a considerable number of measurements have
been performed the device begins to issue continuing beep sounds. At the
same time the message Cleaning needed appears on the display. Please
carry out the following procedure:
Fill up a sample cup with clean warm water (45 – 50°С) and place
1.
it on the sample shelf under the pipette.
2.
Press "clean" to initiate the flushing procedure. The device
automatically takes in the necessary amount of water and then
returns it back into the cup. This is repeated 5 consecutive times
and after that the cup has to be removed from the sample shelf.
After the end of the procedure the device is ready for measurement.