Cleaning And Maintenance; Routine Cleaning - Milkotester Master Pro Operating Instructions Manual

Ultrasonic milk analyzer
Table of Contents

Advertisement

CHAPTER SEVEN
7

Routine cleaning

Complete flushing
Weekly cleaning
Power OFF
For high accuracy and reliability of the device, the following cleaning
procedures should be performed on due times.
7.1. Routine cleaning
When the interval between 2 consecutive measurement analyses is
more than 30 minutes or a considerable number of measurements have
been performed the device begins to issue continuing beep sounds. At the
same time the message Cleaning needed appears on the display. Please
carry out the following procedure:
Fill up a sample cup with clean warm water (45 – 50°С) and place
1.
it on the sample shelf under the pipette.
2.
Press "clean" to initiate the flushing procedure. The device
automatically takes in the necessary amount of water and then
returns it back into the cup. This is repeated 5 consecutive times
and after that the cup has to be removed from the sample shelf.
After the end of the procedure the device is ready for measurement.

Cleaning and maintenance

IN THIS CHAPTER
43
Ultrasonic milk analyzers
Operating Instructions Master Pro
Milkotester Ltd

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Master Pro and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents