ULTRASONIC PORTABLE MILK ANALYZER
CHAPTER FIVE
IN THIS CHAPTER
Routine cleaning
Complete flushing
Weekly cleaning
For high accuracy and reliability of the device, the following cleaning
procedures should be performed on due times.
5. 1. Routine cleaning
When the interval between 2 consecutive measurement analyses is
more than 30 minutes or a considerable number of measurements have
been performed the device begins to issue continuing beep sounds. At
the same time the message Cleaning needed appears on the display.
Please carry out the following procedure:
1.
Fill up a sample cup with clean warm water (45 – 60°С) and
place it on the sample shelf under the pipette.
2.
Press "EXIT" and choose "CLEAN" from the menu to initiate the
flushing procedure. Then turn the knob to the red point and
back to the black point. This is repeated at least 8 consecutive
times and after that the cup has to be removed from the sample
shelf.
Cleaning and maintenance
- 21 -
w w w . m i l k o t e s t e r . c o m