And then
By performing the procedure described above enter the known
values of the parameters of the milk with low percentage of Fat.
After confirming the last figure of the last parameter by pressing ENTER
the following text appears on the display:
The sample of milk with high Fat must be measured 5 consecutive
times. Put the first cup with the sample and press ENTER. After
measurement the following text appears on the display:
Follow the procedure until the 5
last measurement the following text appears on the display:
The sample of milk with low Fat must be measured 5 consecutive times.
Put the first cup with the sample and press ENTER. After measurement
the following text appears on the display:
Set Low sample
PRO=x.xx
LAC=x.xx
SAL=x.xx
Put sample
High Milk
1/5 times
and Press ENTER
Put Sample
High Milk
2/5 times
and Press ENTER
th
measurement. After performing the
Put Sample
Low Milk
1/5 times
and Press ENTER
Put Sample
Low Milk
2/5 times
and Press ENTER
29
Ultrasonic milk analyzers
Operating Instructions Master Pro
Milkotester Ltd