Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Measurement of pH General information PH probe is a unit, measuring the solution acidity or alkalinity degree. It is measured on scale of 0 to 14. The term pH is derived from "p", the mathematical symbol for the negative logarithm, and "H", the chemical symbol of Hydrogen.
Page 2
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Storage For best results, always keep the pH bulb wet. An optimal storage solution for combination electrode is pH 4 buffer with 225 grams of KCl per liter. Table salt, NaCl, can be used if KCl is not really available.
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master New electrolyte preparation - Open the KCl container; - Add in de-ionized water until it reaches the level of 20 ml; - Close the container and shake it to dissolve the KCl; - Add in fresh electrolyte until it reaches the level of the refilling port.
Page 4
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master - Clogged Reference Junction: pH electrodes have junction, which allows the internal fill solution of the measuring electrode to leak out into the solution being measured. The junction can become clogged by contamination in the solution.
Page 5
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Temperature effects on the value of pH buffers T (ºC) 4.01 7.12 9.52 4.00 7.06 9.38 4.00 7.02 9.26 4.01 7.00 9.21 4.01 6.99 9.16 4.03 6.97 9.06 4.06 6.97 8.99 4.10 6.98...
Page 6
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Remove the protective cup of the electrode. Dip the electrode into the low value buffer solution and make sure that the junction zone of the electrode is at least at 30 mm below the surface of the buffer solution and leave it at rest.
Page 7
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Setting Ref. Value: xxxxx Wait until the electrode and the buffer stabilize /the reference value should change the least possible/. Press ENTER to set the value Please note: The pH electrode must be calibrated on a regular base in order to obtain high accuracy results of the measurement.
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Auto pH measure Manual pH measure Under the manual service menu, the milk sample is put under the pH probe, then you press the pH button and the measurement will last without interruption (until the pH button is pressed again).
Page 9
Appendices towards the Operation Manual of Ultrasonic milk analyzer Master Now you can start measuring the other components of milk. MENU > Settings > Set Thermometer Press ENTER and the following text will appear on the display: Put probe 15C and press ENTER Put a sample of water with temperature 15...