Platform Mounting Plate; Platform Maintenance; Section 6 Maintenance - Thermo Scientific MaxQ SHKA4000 Operating Manual

Maxq 4000 incubated and refrigerated shakers shke4000 and shka4000
Hide thumbs Also See for MaxQ SHKA4000:
Table of Contents

Advertisement

Section 6
Maintenance
Replacing Platform
Mounting Plate
Platform
Maintenance
6-2
Incubated and Refrigerated Shakers
Note While adjusting the platform retaining screws, make certain lid
switch, located at back of chamber, is activated (pressed in).
(cont.)
3. While it is slowly orbiting, slightly tighten each of the A screws in
Figure 1. Repeat this procedure by slightly tightening all the B screws,
then complete procedure by tightening the C screws.
Caution All screws must first be slightly tightened to seat plate properly.
4. Repeat the screw-tightening procedure at 50-100 RPM until all 9
screws are securely tightened.
Figure 6-1. Platform Mounting Plate
Suggested with every 3 months of constant use.
Caution Any internal adjustments or repairs must be performed by a
qualified service representative.
Remove the platform by loosening 4, thumbscrews in the platform center.
Remove the sheet metal panel (9 screws) under the platform to expose the
belt and interior parts. Inspect the drive belt for wear. Order a replacement
if necessary.