Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Weight
Enclosure rating
Protection class
Pollution degree/installation category
Power connector via external power
supply
Relative humidity
Operating temperature
Storage temperature
Measurement method
Display
Regulatory compliance
Laser
Optical light source
Measurement modes
Range
Stray light
Accuracy
Linearity
Repeatability
Signal average time
Detection limit
Calibration options
Verification options
Verification (Link2SC)
Details
41 x 28 x 12.5 cm (16 x 11 x 7.7 in.)
2.37 kg (5.23 lb)
IP20
III for the instrument
II for the external power supply
2, II
Input: 100–240 V, 50/60 Hz
Output: 15 VDC, 2A
5 to 95% at different temperatures, non-condensing
10 to 40 °C (50 to 104 °F)
–30 to 60 °C (–22 to 140 °F)
90-degree nephelometry, 360 degrees around the axis of the incident light
beam
17.8 mm (7 in.) color touch screen
DIN EN ISO 7027
Class 1 laser product: contains a non-user serviceable laser
850 nm, maximum 0.79 mW
NTU, FNU, TE/F, FTU, EBC, mNTU or mFNU
0 to 1000 FNU
<0.010 FNU
+/- 2 % of reading plus 0.01 FNU from 0 to 40 FNU
+/- 10 % of reading from 40 to 1000 FNU
Better than 1 % 0 to 40 NTU on formazin at 25 °C (77 °F). Allows for
accurate calibration at 20 FNU
± 1% of the reading or ± 0.01 FNU, whichever value is more
Off, 5 to 90 seconds (default=30 seconds)
0.0003 FNU
®
StablCal
: at 20 FNU from 0 to 40 FNU; at 20 FNU and 600 FNU for full
range
User prepared formazin, SDVB, kaolin or EBC via factory-defined or
custom curves
®
StablCal
, formazin (0.1 to 40 NTU), Hach solid verification standards
Verification of the measurement value to compare values to a online
instrument via Link2SC.