Data Management; Configure The Data Log; Maintenance - Hach TU5200 User Manual

Hide thumbs Also See for TU5200:
Table of Contents

Advertisement

The instrument is factory calibrated. Do calibration procedures as necessary. Do the steps that follow
to do a full range calibration procedure.
1. Clean the vial, then carefully invert the vial. Do not cause air bubbles. Refer to
verification or calibration
2. For vials with an RFID tag, scan the vial in front of the RFID scanner. Refer to
scanner (optional)
3. Push Calibration and follow the instructions on the display.
4. Put the vial in the vial compartment. Refer to
5. Push Read to start the calibration.
6. Wait until the procedure is complete, then open the lid to remove the vial. The procedure is
complete when the value and range shows on the display.
7. Remove the first calibration vial, then do steps
8. Read the display for the calibration information.
9. Push Store to complete the calibration and keep the results.

Data management

Configure the data log

The data log keeps all saved comments and readings from measurements, calibrations and
verifications. Use the steps that follow to select, manage and store data.
1. Push Data Log and select a data log to show.
2. Select an option.
Option
Description
Filter
Sets the data filtering to on or off. If on, select and arrange data by time interval or operator ID.
View details Shows information about a selected data log. Push the comments icon to enter any necessary
information.
Options
Shows the current options (contextual). View Graph—See data as a graph, set control limits to
on or off (0.800 = upper control limit or 0.200 = lower control limit). Add data—Select a sample
ID to add to the graph. Curve style(???WHAT IS THE PROPER NOUN???)—Selects the
curve style as crosses, rectangular dots or a line. Delete—Removes selected data from the
data log. Send data—Sends selected data to a connected USB flash drive or PC.

Maintenance

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and
the user risks injury if exposed to the laser.
on page 18.
on page 17.
-BETA DRAFT-
Put the vial into the instrument
1
to
5
again for the second calibration vial.
C A U T I O N
C A U T I O N
C A U T I O N
Prepare for a
Use the RFID
on page 15.
English 19