10.4 Command List
The table below describes the available commands via RS485 or LAN/Ethernet interface.
Description
Burst string format
Show list of commands
Measured attenuation
Burst speed
Baud rate (6)
Digital filter
DHCP (Dyn. Host Config. Protocol)
Emissivity
Error Codes (9)
Valley hold time (4)
Average time (4)
Gateway Address
Top of mA temperature range
Sensor internal ambient
IP Address
Switch panel lock
Relay alarm output control
Bottom of mA temperature range
Mode–ER series
MAC Hardware Address
Target temp – 1-color narrow
Net Mask
Output current
ENDURANCE 1R
info@Raytek-Direct.ca
Char
Format
P
B
(2)
(1)
(1)
$
(3)
√
?
√
B
nn - nnn
√
√
BS
n - nnnnn
√
D
nnn - nnnn
√
√
√
√
√
√
√
DF
n
√
DHCP
n
√
E
n.nn
√
√
EC
nnnnnnnn
√
F
n.n - nnn.n
√
√
G
n.n - nnn.n
√
√
GW
nnn.nnn.nnn.nnn √
H
n.n – nnnn.n
√
√
I
n.n - nnn.n
√
√
IP
nnn.nnn.nnn.nnn √
J
X
√
K
n
√
L
n.n – nnnn.n
√
√
M
n
√
√
MAC
nnnnnnnnnnnn
√
N
n.n - nnnn.n
√
√
NM
nnn.nnn.nnn.nnn √
O
nn
√
Rev. E1 Sep 2015
Programming Guide
S
Legal Values
(1)
√
(3)
00 to 100%
√
5 msec – 10 sec (5 – 10000)
√
012 = 1200
baud
√
024 = 2400
baud
√
096 = 9600
baud
√
192 = 19200 baud
√
384 = 38400 baud
√
576 = 57600 baud
√
1152 = 115200 baud
√
0 = OFF, 1 = ON
√
0 = OFF, 1 = ON
√
0.10 – 1.10
0000 – FFFF (Hex)
√
0.0 – 300.0 sec (300 s = ∞)
√
0.0 – 300.0 sec (300 s = ∞)
√
0.0.0.0 - 255.255.255.255
√
0.0 – 9999.0 (°C or °F)
0.0 – 999.0 (°C or °F)
√
0.0.0.1 - 255.255.255.255
√
L = Locked
U = Unlocked
√
0 = off
1 = on
2 = Normally Open
3 = Normally Closed
√
0.0 – 9999.0 (°C or °F)
√
1 = 1 - color
2 = 2 - color
e.g. 001d8d2aaa01
√
0.0.0.1 - 255.255.255.255
√
00 = controlled by unit
02 = under range
21 = over range
00 – 20 = current in mA
Factory Default
UTSI
50 msec
38400 baud
0
0
1.00
0.0
0.0
192.168.42.1
Upper end of sensor range
192.168.42.132
Unlocked
2
Lower end of sensor range
2
Set at factory calibration
255.255.255.0
00
51
1.800.868.7495