Options
Air flow should be 1.4 to 2.5 l/sec at 25°C (77°F). Water flow should be approximately 1.0 to 2.0 l/min
(water temperature between 10 and 27°C / 50 to 80.6°F). The maximal pressure limit is 5 bar (73 PSI).
Chilled water below 10°C (50°F) is not recommended.
The Air/Water Cooled Housing is delivered with plugs only removable with a 5 mm
hex wrench. Check your supplier for appropriate fittings.
Figure 27: Endurance Head with Air/Water-Cooled Housing Option
For ambient temperatures exceeding 175°C (350°F), the ThermoJacket can be used. This
accessory allows operation at ambient temperatures up to 315°C (600°F)!