Sper scientific 800005 Manual

Sper scientific 800005 Manual

Type k/j
Hide thumbs Also See for 800005:

Advertisement

Type K/J
Thermometers
800005

Advertisement

Table of Contents
loading

Summary of Contents for Sper scientific 800005

  • Page 1 Type K/J Thermometers 800005...
  • Page 2: Table Of Contents

    CONTENTS INTRODUCTION . . . . . . . . . . . . . . . . . . . 2 FRONT PANEL DESCRIPTION .
  • Page 3: Front Panel Description

    FRONT PANEL DESCRIPTION 1 . K / J type temperature sensor connector 2 . LCD display 3 . ON/OFF button 4 . HOLD button 5 . Timer function control button 6 . Relative readout button 7 . MAX/MIN Average control button 8 .
  • Page 4: Lcd Display

    LCD DISPLAY Main Display Timer Display This indicates that the minus temperature is sensed. It indicates that the timer is expressing in hours and minutes. Centigrade and Fahrenheit indication. °C °F Thermocouple Type Indication This indicates that the display data is being hold. HOLD The Maximum value is now being displayed The Minimum value is now being displayed...
  • Page 5: Operation Instruction

    OPERATION INSTRUCTIONS Power-Up Press the POWER button to turn the thermometer On or OFF Connection the Thermocouples For measurement, plug the thermocouple into the input connectors . Selecting the Temperature Scale When the meter was first power on, the default scale setting is set at Celsius (°C) scale .
  • Page 6: Data-Hold Operation

    Data-Hold Operation The user may hold the present reading and keep it on the display by pressing the HOLD button . When the hold data is no longer needed, one may release the data- hold operation by pressing HOLD button again .
  • Page 7: Max/Min/Avg Operation

    MAX/MIN/AVG Operation When one press the MAX/MIN/AVG button the meter will enter the MAX/MIN mode . Under this mode the maximum value, minimum value and average value of latest 4 data is kept in the memory simultaneously and updated with every new data . When the MAX symbol is display, the Maximum is shown on the display .
  • Page 8: Auto Power Off

    Auto Power Off By default, when the meter is powered on, it is under auto power off mode. The meter will power itself off after 30 minutes if no key operation and RS232 communication . Key combination at power on or RS232 communication can disable auto power off.
  • Page 9 Resolution K-Type: -200°C to 800°C 0 .1°C; 800°C to 1370°C 1°C -328°F to 1000°F 0 .1°F; else 1°F J-Type: -200°C to 600°C 0 .1°C; 600°C to 760°C 1°C -328°F to 1000°F 0 .1°F; else 1°F Maximum Voltage at Thermocouple Input 60V DC, or 24 Vrms AC Environmental •...
  • Page 10 Accuracy at ( 23 ± 5°C ) Range Accuracy K-Type -200°C to 1370°C ±(0 .1% reading + 0 .7°C) J-Type -200°C to 760°C ±(0 .1% reading + 0 .7°C) K-Type -328°F to 2498°F ±(0 .1% reading + 1 .4°F) J-Type -328°F to 1400°F ±(0 .1% reading + 1 .4°F) Temperature Coefficient: For ambient temperatures from 0°C to 18°C...
  • Page 11 Sample Rate 3 .3 times per second Dimension 184×64×30 mm Weight 210 g Approx . Accessory K Type Bead Probe, Battery, Carrying Case, Instruction Menu . Option Soft Ware Package (Program, RS232 Connection Cable), AC Adapter Power requirement 9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22 Battery Life Approx .
  • Page 12: Warranty

    WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for five (5) years from the date of purchase, and agrees to repair or replace any defective unit without charge . If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available .

Table of Contents