4-channel datalogging thermometer
Hide thumbs Also See for 800024:

Advertisement

4-Channel
Datalogging
Thermometer
800024

Advertisement

loading

Summary of Contents for Sper scientific 800024

  • Page 1 4-Channel Datalogging Thermometer 800024...
  • Page 2 4-Channel Datalogging Thermometer - 800024 Copyright ©2015 by Sper Scientific ALL RIGHTS RESERVED. Printed in the USA The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may...
  • Page 3 TABLE OF CONTENTS 1. INTRODUCTION ....3 2. SYMBOL DEFINITION AND BUTTON LOCATION . . 4 3. OPERATION INSTRUCTIONS ... . 6 4.
  • Page 4: Symbol Definition And Button Location

    SYMBOL DEFINITION AND BUTTON LOCATION 11 10 9 1. Thermocouple Type Indication. 2. It indicates the value below is T1-T2 sensor. 3. The Battery is not sufficient for proper operation. 4. This indicates that the minus temperature is sensed. 5. This indicates auto power off is enabled. 6.
  • Page 5 BUTTON LOCATION: 1. K type temperature sensor 8. Offset calibration screw (on side) T1 to T4 input connector 9. Digital output connector (on side) 2. LCD display 10. MAX MIN function control button 3. ON/OFF & Backlight button 11. °C, °F control button 4.
  • Page 6: Operation Instructions

    OPERATION INSTRUCTIONS 1. Power-Up & Turn ON/OFF backlight The POWER key turns the Thermometer ON or OFF and backlight ON & OFF. Press it once to turn on the Thermometer. Press it again for moment to turn ON or OFF backlight. Press and hold this button 3 second to turn OFF the power.
  • Page 7 mode, there will be two continuous beeps) 5. T1-T2 Operation: When this button is pushed, 1 – 2 will be shown on the upper right hand side LCD display to indicate that the tester is under T1 minus T2 mode. The temperature difference is shown on the right hand side display as shown in Fig.
  • Page 8 button to cancel. 8. Recording Interval Setup: 1. Press and hold T1-T2 button and then power on the meter: 2. Press MAXMIN(INTV) 3. Press REC • or °C/°F • to increase or decrease number, press MAXMIN (INTV) to adjust next item, then press MAXMIN (INTV) to finish.
  • Page 9 To exit the MAX/MIN mode, one may press and hold MAX MIN for two seconds. 10. Auto Power Off: By default, when the meter is powered on, it is under auto power off mode. The meter will power itself off after 30 minutes if no key operation and no RS232 communication combination at power on can disable auto power off.
  • Page 10 1000°C ± 1°C 1900°F ± 1°F 13. Digital Output: The Digital Output is a 9600 bps N 81 serial interface. The RX is a 5V normal high input port. The TX is a 5V normal high output port. SETUP TESTLINK SE-309 - RS323 INTERFACE SOFTWARE: ̀...
  • Page 11 • The TestLink package contains: 1. 80mm CD. 2. Custom designed RS232 cable for TestLink. ̀ • System Required: Windows 95 or Windows 98 or Windows NT 4.0 above • Minimum Hardware Required: PC or NoteBook with Pentium 90MHz or higher, 32 MB RAM ;...
  • Page 12: Specifications

    At least 5 MB byte hard disk space available to install TestLink. Recommended resolution 800X600. ̀ • Install TestLink: 1. We recommend close all other application before installing TestLink. 2. Insert setup CD disk to CD disk drive. 3. Choose the Start button on the Taskbar and select Run. 4.
  • Page 13 Resolution: -200°C ~ 200°C 0.1°C; 200°C ~ 1370°C 1°C -200°F ~ 200°F 0.1°F; else 1°F Input Protection at Thermocouple Input: 60V DC, or 24 Vrms AC Environmental: • Operating Temperature and Humidity: 0°C ~ 50°C (32°F ~ 122°F) ; 0 ~ 80% RH •...
  • Page 14 200°F ~ 2498°F ±(0.3% reading + 2°F) Temperature Coefficient: For ambient temperatures from 0°C ~ 18°C and 28°C ~ 50°C, for each °C ambient below 18°C or above 28°C add the following tolerance into the accuracy spec. 0.01% of reading + 0.03°C (0.01% of reading + 0.06°F) Note...
  • Page 15 Electromagnetic Compatibility: Total accuracy = specified accuracy ± 2°C (3.6°F) Sample Rate: 3 seconds per period Dimension: 7¼" × 2½" × 1¼" (184 × 64 × 32 mm) Weight: 8.4 oz (238 g) 0090 Power requirement: 9 Volt Battery Battery Life: Approx. 100 hrs with alkaline battery AC Adapter: 9V DC ±15% 100 mA Plug Diameter: 3.5 mm ×...
  • Page 16: Warranty

    If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available. This warranty does not cover probes, batteries, battery leakage, or damage resulting from accident, tampering, misuse, or abuse of the product.