INTRODUCTION Your digital thermometer accurately measures temperature in Centigrade or Fahrenheit from -200°C to 1370°C (-328°F to 2498°F). It conforms to the National Bureau of Standards and IEC584 temperature and voltage tables for K-type thermocouple probes. Features include a data hold function, maximum and minimum value recording, memory and clock functions, and a low battery indicator.
MEASUREMENT PROCEDURE Setting the Clock 1. Press the green on/off button and MAX/MIN button simultaneously, release. 2. The meter will show SEL. 3. Press the CLOCK button to select the year, month, day, hour and minute (in that order). 4. To set the year, month, day, hour and minute, press the ▲ button to increase the number and the ▼...
Clearing Data from Memory 1. With the meter turned off, press and hold both the POWER button and the REC button for at least 2 seconds. 2. Release the power button, then release the REC button. CLR will be displayed. Relative Operation When the meter is in relative operation mode the °C/°F button wil be disabled.
AUTOMATIC SHUTOFF If the meter is not used for 30 minutes, it will shut off automatically. To disable automatic shutoff: 1. With the meter turned off, press and hold down the HOLD button. 2. Press the POWER button to turn the meter on. Release the buttons.
Page 7
Command K: Return 4 bytes. For example, when sending command “K” to meter, it will return “3”,”0”,”5”, ASCII(13). Command U: Return 32768 bytes. Command P: Instead of returning all 32768 bytes, it only returns recorded data. Command H: Equivalent to one pushing on the HOLD button and no message is returned.
Page 8
3rd BYTE: bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0 Auto Power Off Memory Full Resolution Sign bit0: 1→main window value is OL, 0→not OL bit1: 1→main window value is minus, 0→main window value is plus. bit2: 1→4th byte and 5th byte represent #### , 0→4th byte and 5th byte represent ###.# bit6: 1→Memory is full.
Page 9
Software Download: With technology upgrading, some computers no longer have a disc drive for you to install the software that comes with your meter. You can download the software that came with your meter directly. Go to the www.sperdirect.com/software.htm and find your meter or search for you meter (800008) and download from product page.
Page 10
Main Menu: Link Test: Open Link Test window to search for thermometer connected to PC. When you start the THERMOLINK, this window will display first and search for thermometer. The result will be shown in the text box. Control Panel: By opening the Control Panel Window, the user can control thermometer via the button in this window.
WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for a period of five (5) years from the date of purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available.