Emerson Rosemount 400 Instruction Manual page 20

Contacting conductivity sensors
Table of Contents

Advertisement

Troubleshooting
May 2017
4.1.1
Checking the temperature element
Disconnect leads and measure resistance shown. The measured resistance should be close to the
value in the table.
Figure 4-1 Checking the temperature element
4.1.2
Checking the continuity and leakage
Disconnect electrode leads and measure resistance and continuity as shown. Sensor must be dry
when checking resistance between electrode leads.
Figure 4-2 Checking the continuity and leakage
14
Instruction Manual
Temperature
Resistance in ohms
(°C)
Pt 100
Pt 1000
0
100.0
1000
10
103.9
1039
20
107.8
1078
30
111.7
1117
40
115.5
1155
50
119.4
1194
LIQ-MAN-400
Troubleshooting

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 400vp

Table of Contents