Emerson Rosemount 400 Instruction Manual page 13

Contacting conductivity sensors
Table of Contents

Advertisement

Instruction Manual
LIQ-MAN-400
2.2.1
Wiring through junction box
Rosemount 400 Contacting Conductivity Sensors can have an optional integral junction
box mounted on the end of the sensor. See
If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point.
Use cable 23747-00 (factory-terminated) or 9200275 (raw cable).
Figure 2-13 Sensor-mounted junction box wiring
Notes:
1. The gray sensor wire is connected to the junction box, which makes electrical contact with the
OUTER electrode.
2. Terminals in junction box are not numbered. Refer to transmitter wiring diagram for connections
at transmitter.
Installation
Figure 2-13
for wiring instructions.
Installation
May 2017
7