(8) [Auto discharge function]
This function allows electric charges stored in the capacitance of the
circuit under test to be automatically discharged after measurement.
Set the TEST button or remote control switch to off with the test leads
connected. Discharge can be monitored by the readings displayed at the
lower left of the LCD and also by live circuit warning LED, red backlight
and blinking
(9) Press the POWER button to power off the instrument when measurement
completes, and then disconnect the test leads from the instrument.
● Never touch the circuit under test immediately after measurement.
Capacitances stored in the circuit may cause electric shock. Leave the
test leads connected to the circuit and do not touch the circuit until the
voltage value displayed at the lower left corner of the LCD becomes
positive dc value, live circuit warning LED turns off and audible warning
stops.
.800.561.8187
mark.
Blink
Voltage
DANGER
- 30 -
̶ 30 ̶
www.
.com
Warning example
Fig. 7-6