Voltage/ Earth voltage measurement
Measuring range
±5 to ±1000 V DC
Measuring method: True RMS
* Automatically detect AC/ DC when an input voltage is 5 V or higher and
show AC or DC symbol on the LCD.
* Input terminals are shown in the following table.
LINE-EARTH
C(H)-E
P(S)-E
Earth resistance measurement
Precision/
simplified
Measuring method: Constant current inverter/ 825 Hz
* For precision measurement, auxiliary earth resistance should be 100 Ω
±5% or less.
.800.561.8187
5 to 600 V AC
(45 - 65Hz)
when selecting a voltage function
when selecting a simplified measurement function
when selecting a precision measurement function
Range
(3-range auto)
20Ω range
200Ω range
2000Ω range
www.
Display range (2-range auto)
300V range: 0.0 to 314.9 V
600V range: 240 to 629 V
500V range: 0.0 to ±524.9 V
1000V range: ±400 to ±1049 V
Measuring
range
0.00 - 2000 Ω
20Ω range
: approx. 3 mA
200Ω range : approx.2 mA
2000Ω range : approx. 1 mA
- 10 -
̶ 10 ̶
.com
Accuracy
±1%rdg±4dgt
Display
Accuracy
range
0.00 - 20.99 Ω
±3%rdg±0.1 Ω
16.0 - 209.9 Ω
±3%rdg±3dgt
160 - 2099 Ω
information@itm.com