Service - PIN
Info
VEGAPULS 66 • 4 ... 20 mA/HART - two-wire
6 Set up with the display and adjustment module PLICSCOM
Inner diameter of the standpipe (with standpipe versions)
Vessel form
Damping
Linearisation curve
Sensor-TAG
Displayed value
Display unit
Scaling
Current output
Unit of measurement
Language
The following safety-relevant data are not read out or written:
HART mode
PIN
SIL
Copy sensor data
Copy sensor data?
In this menu item, the PIN is activated/deactivated permanently.
Entering a 4-digit PIN protects the sensor data against unauthorized
access and unintentional modifications. If the PIN is activated perma-
nently, it can be deactivated temporarily (i.e. for approx. 60 min.) in
any menu item. The instrument is delivered with the PIN set to 0000.
PIN
Disable permanently?
Only the following functions are permitted with activated PIN:
Select menu items and show data
Read data from sensor into the display and adjustment module.
In this menu item the most important sensor information can be
displayed:
Instrument type
Serial number: 8-digit number, e.g. 12345678
Instrument type
Serial number
Date of manufacture: Date of the factory calibration
Software version: Edition of the sensor software