Installation of a Dräger Polytron Pulsar 2

Mounting the unit

The Dräger Polytron Pulsar 2 should be mounted to a stable structure free of excessive vibration. Good choices would be a steel bulk-
head, brick wall, concrete lamppost or a rigid steel structure. Avoid flimsy metal structures that may flex, or wooden structures that may
warp. In open areas a suitable structure close to the ground would consist of a five inch nominal (141mm outside diameter) steel pipe
driven 1 metre into firm ground or embedded into a concrete foundation. Tall structures should be suitably guyed or braced.
Figure 1: Wall mounting of Dräger Polytron Pulsar 2 with EEx e junction box
FRONT VIEW
Front cover fitted
Rear cover
216
Electrical connection via M20 certified
Ex e cable glands - not supplied.
For connection options refer to figure 4, 5 and 6.
Figure 2: Wall mounting of Dräger Polytron Pulsar 2 with eX link plug and socket
FRONT VIEW
Rear cover
Front cover fitted
216
For opposite hand mounting (see figure 1 and 2).
Remove front cover annd screws (if supplied).
Remove Gimbal Clamps.
Reverse the Detector taking care not to kink cable and refit the Gimbal Clamps.
Refit the cover.
10
Gimbal Clamps
Locking screws for
horizontal adjustment.
Locking screws for
vertical adjustment.
Mounting bolts M8 max.
Terminal Box cover bolts.
DO NOT OPEN WHEN ENERGISED
Terminal Box fixing bolts
Gimbal Clamps
Locking screws for
horizontal adjustment.
Locking screws for
vertical adjustment.
Mounting bolts M8 max.
SIDE VIEW
338
Front cover removed
71
SIDE VIEW
338
Front cover removed
71
eX Link plug
eX Link socket/coupler
Minimum bend diameter
= 6 x cable diameter
Factory fitted Ex e
certified cable gland
for Power/Signal cable
Front cover
fixing screws