Introduction; Required Equipment - Tektronix DMM4020 Technical Reference Manual

Hide thumbs Also See for DMM4020:
Table of Contents

Advertisement

Introduction

This chapter of the Calibration Manual provides performance tests to verify the Meter is
operating within published specifications as well as a complete calibration procedure.
The performance test and, if necessary, the calibration procedure can be performed both
periodically and after service or repair. A one-year calibration interval is recommended to
maintain the accuracy of the Meter.
The performance tests can be used as an acceptance test upon receipt of the Meter. Use
the 90-day specifications when performing an acceptance test after performing a
calibration.

Required Equipment

Table 3-1 lists the equipment required for performance testing and calibration of the
Meter.
Function
Instrument Type
Volts DC
Calibrator
4-wire short
Volts AC
Calibrator
Amplifier
Resistance
Calibrator
Amps DC
Calibrator
Amplifier
4-wire short
Amps AC
Calibrator
Amplifier
Frequency
Function Generator
Resistance
Calibrator
Adjust
Table 3-1. Required Test Equipment
Model
Fluke 5700A
Tektronix low thermal 4-
wire short or equivalent
Fluke 5700A
Fluke 5725A
Fluke 5700A
Fluke 5700A
Fluke 5725A
Tektronix low thermal 4-
wire short or equivalent
Fluke 5700A
Fluke 5725A
Fluke 271 or equivalent
Fluke 5520A
Performance Verification and Calibration
Comments
Figure 3-1
Tektronix PN 013-0369-00
Figure 3-1
Figures 3-2 & 3-3
Figure 3-4
Figure 3-5
Tektronix PN 013-0369-00
Figure 3-4
Figure 3-5
Requires frequency specification
better than 0.0025 %
Requires resistance accuracy better
than 0.0045 %
3

Introduction

3-3