Tektronix DMM4020 Technical Reference Manual page 48

Hide thumbs Also See for DMM4020:
Table of Contents

Advertisement

DMM4020
Technical Reference
Step
Func./Command
96
AAC
97
AAC
98
AAC
99
AAC
Current DC: zero points – Adjust with 5700A in standby or disconnected
100
ADC
Current DC: Gains – Adjust using 5700A/5725A
101
ADC
102
ADC
103
ADC
104
ADC
105
ADC
106
ADC
107
ADC
108
ADC
109
ADC
110
ADC
111
ADC
112
ADC
113
ADC
114
ADC
115
ADC
116
ADC
117
ADC
118
ADC
119
ADC
120
ADC
121
ADC
122
ADC
123
ADC
124
ADC
When the calibration point finishes calibration and another calibration point is part of the
selected function, a new recommended calibration value will be displayed automatically.
Input the proper signal and repeat step 2 above.
3-16
Table 3-9. Adjustment Steps (cont.)
Point
Input Signal
0.5 A
0.5 A@500 Hz
2 A
2 A@500 Hz
5 A
5 A@500 Hz
10 A
10 A@500 Hz
0 µA/0 mA/0 A
Open, no connection
-199.9 µA
-199.9 µA
-100 µA
-100 µA
100 µA
100 µA
199.9 µA
199.9 µA
-1999 µA
-1999 µA
-1000 µA
-1000 µA
1000 µA
1000 µA
1999 µA
1999 µA
-19.99 mA
-19.99 mA
-10 mA
-10 mA
10 mA
10 mA
19.99 mA
19.99 mA
-199.9 mA
-199.9 mA
-100 mA
-100 mA
100 mA
100 mA
199.9 mA
199.9 mA
-1.999 A
-1.999 A
-1 A
-1 A
1 A
1 A
1.999 A
1.999 A
-10 A
-10 A
-5 A
-5 A
5 A
5 A
10 A
10 A
Description
Gain of AAC, 10 A range
Zero point of ADC, all ranges
Gain of ADC, 200 µA range,
need characterize 5520A with
8508A
Gain of ADC, 2000 µA range,
need characterize 5520A with
8508A
Gain of ADC, 20 mA range
Gain of ADC, 200 mA range,
need characterize 5520A with
8508A
Gain of ADC, 2 A range, need
characterize 5520A with
8508A@±1.999 A
Gain of ADC, 10 A range