Alternate Current Verification
To verify the Amps AC function of the Meter, connect it to the test equipment as shown
in either Figure 3-4 or Figure 3-5 depending on current level and apply the current levels
listed in Table 3-7.
Nominal Input
Ampl.
Freq.
1 mA
45 Hz
19 mA
20 Hz
19 mA
45 Hz
19 mA
2 kHz
10 mA
45 Hz
190 mA
20 Hz
190 mA
45 Hz
190 mA
2 kHz
0.1 A
45 Hz
1.9 A
20 Hz
1.9 A
45 Hz
1.9 A
2 kHz
0.5 A
45 Hz
10 A
40 Hz
10 A
45 A
10 A
1 kHz
Table 3-7. Alternating Current Verification Steps
90-Day Test Limit
Range
High
20 mA
1.0125 mA
20 mA
19.2000 mA
20 mA
19.058 mA
20 mA
19.0575 mA
200 mA
10.125 mA
200 mA
191.620 mA
200 mA
190.575 mA
200 mA
190.575 mA
2 A
0.10125 A
2 A
1.92000 A
2 A
1.90575 A
2 A
1.90575 A
10 A
0.5118 A
10 A
10.1100 A
10 A
10.0450 A
10 A
10.0450 A
Performance Verification and Calibration
Alternate Current Verification
1-Year Test Limit
Low
High
0.9875 mA
1.0150 mA
18.8000 mA
19.2495 mA
18.943 mA
19.0690 mA
18.9425 mA
19.0690 mA
9.875 mA
10.150 mA
188.380 mA
192.020 mA
189.425 mA
190.690 mA
189.425 mA
190.690 mA
0.09875 A
0.10150 A
1.88000 A
1.92495 A
1.89425 A
1.90690 A
1.89425 A
1.90690 A
0.4883 A
0.5145 A
9.8900 A
10.1370 A
9.9550 A
10.0620 A
9.9550 A
10.0620 A
3
Low
0.9850 mA
18.7505 mA
18.9310 mA
18.9310 mA
9.850 mA
187.980 mA
189.310 mA
189.310 mA
0.09850 A
1.87505 A
1.89310 A
1.89310 A
04855 A
9.8630 A
9.9380 A
9.9380 A
3-11