Send the string "EXIT" to exit the calibration mode at any time. If a function has not
been completely calibrated, a CAL ERROR will be displayed.
When the Meter returns a recommended calibration value followed by
"W4", 4-wire compensation is enabled on the input.
When in current calibration, if "Disconnect Test Leads!" is returned, all
test leads should be removed from the Meter.
When "Connect to mA socket!" is returned from the Meter, use the mA
input terminal to connect to the calibrator. When "Connect to 10A socket"
is returned from the Meter, use the 10A input terminal to connect to the
calibrator.

Calibration Points

Listed below are the calibration values required for each calibration point of the
calibration procedure.
{//adc
{0.0, -0.0001999, -0.0001, 0.0001, 0.0001999}, range 1
{0.0, -0.001999, -0.001, 0.001, 0.001999}, range 2
{0.0, -0.01999,
{0.0, -0.1999,
{0.0, -1.999,
{0.0, -10.0,
},
{//aac
{0.0, 0.001, 0.005, 0.01, 0.01999},
{0.0, 0.01, 0.05, 0.1,
{0.0, 0.01, 0.5,
{0.0, 1.0,
},
3-18
Note
Note
Note
-0.01,
0.01,
0.01999},
-0.1,
0.1,
0.1999},
-1.0,
1.0,
1.999},
-5.0,
5.0,
10.0},
0.1999},
1.0,
1.999},
2.0,
5.0,
10.0},
range 3
range 4
range 5
range 6
range 1
range 2
range 3
range 4