Display Tests
The display test consists of turning on all the display elements and checking what appears
in the display with Figure 2-3. To turn on all the display elements:
1. Turn the Meter on by moving the rear-panel power switch to the "On" position.
2. Put the Meter into standby by pressing the front-panel power button until the display
3. While holding , push in on the front-panel power button until the display
4. Check the elements of the display against those elements appearing in Figure 2-3.
5. Push the front-panel power button twice to clear the display test.
2-8
extinguishes.
illuminates.
Figure 2-3. Display Elements
ffw008.eps