Introduction; Calibration Interval; Environmental Conditions; Recommended Test Equipment - Tektronix DM2510 Service Procedure

Hide thumbs Also See for DM2510:
Table of Contents

Advertisement

Introduction
ADJUSTMENT PROCEDURE
Use of this adjustment procedure to
restore DM2510G or DM2510
to the
original performance requirements .
This procedure needs not to be performed
unless the Performance Check Procedure cannot be completed satisfactorily or
the instrument fails to meet the Performance Requirements of the electrical
characteristics
listed
in
the
Specification
section
in
the
DM2510G/2510
instruction manual ( Part No. 070-8479-00).
Calibration Interval
Adjustment should be performed if the performance check procedures show
that the
DM2510G/ DM2510
is
out
of
specification .
Calibration
is
also
recommended following instrument repair or modification .
Environmental Conditions
Calibration
should
be
performed
under
laboratory
conditions
having
an
ambient temperature of 22 - 240C and a relative humidity of less than 70% .
The DM2510G/ 2510 must be turned on and allowed to warm up for at least 30
minutes before beginning the performance checks . After exposure to or storage
in
high
humidity
(condensing)
environment ,
one
hour
warm -up
time
is
required .
Recommended Test Equipment
The Test equipment listed in Table 1 , or equivalent is suggested to perform the
DM2510G/2510 calibration .
Alternate equipment may be used as long as the
accuracy is at least as good as the specifications listed in the table.
DM2510/DM2510G
Page 3 of 17
Adjustment Procedure
May 6 , 1994
Revision A
S

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dm2510g

Table of Contents