SPM Flex Gas Detector

In-line filters

Use an external filter to protect the tubing and the detector from contamination�
Use particulate filter part number 780248 for non-corrosive gases� Use filter
1991-0147 for corrosive gases� Refer to Detectable Gases on page 51 for specific
gases� Filters should be replaced every 3 to 6 months of operation, depending on
the cleanliness of the installation environment�

Optional sampling wand

The sampling wand is connected to the inlet port and used to detect toxic gas
at specific locations
2
entering the unit� All gas-wet surfaces are either Teflon-coated or made of Kynar
to be compatible with sticky corrosive gases� An appropriate filter should be used
for the gas type being monitored�
The sampling wand is supplied with the mounting bracket and required hardware�
filter
wand
bracket
Figure 21. Attaching the sampling wand bracket to the detector
2 Do not use the sampling wand with diisocyanates.
30