Honeywell SPM Flex User Manual page 25

Single point monitor gas detector
Hide thumbs Also See for SPM Flex:
Table of Contents

Advertisement

Tubing (optional)
Sample and exhaust tubing calculations
This table shows the flow rate, tubing length, transport time, and maximum
pressure and vacuum at the inlet and exhaust points�
Tubing lengths vary among gases� If the pressure or vacuum on the inlet/exhaust
lines does not meet the recommended values, the detector may encounter flow
faults� The maximum total tubing length (inlet + outlet) is 100 ft� (30 m)�
Description
Tubing length, ft. (m)
Transport time (sec)
Flow rate (cc/min.)
Inlet
Tubing OD. in. (mm)
Tubing ID, in. (mm)
Tubing length, ft. (m)
Outlet
Tubing OD, in. (mm)
Tubing ID, in. (mm)
The overall maximum load on the pump between the inlet and the exhaust should not exceed 10 inches H
NOTE
Use Teflon Fluorinated Ethylene Polymer (FEP) tubing to ensure proper sample
transport�
An external filter must be installed on the detector's "in" line, either at the
detector, for local monitoring, or at the end of the sampling line, for remote
monitoring� A filter can also be installed on the exhaust port to reduce noise�
filter
(arrow indicates
direction of flow)
Push
fittings
Figure 16. External filter
Sample Specifications
Value
100 (30)
66 (20)
19
13
700-1200 (flow is set and controlled per calibration)
0.25 (6.35)
0.125 (3.18)
100 (30)
0.25 (6.35)
0.188 (4.76)
SPM Flex Gas Detector
33 (10)
0
7
1
O.
2
25