Loading A Chemcassette Cartridge - Honeywell SPM Flex User Manual

Single point monitor gas detector
Hide thumbs Also See for SPM Flex:
Table of Contents

Advertisement

Loading a Chemcassette cartridge

Close the Chemcassette cartridge door before putting the detector into Monitor
mode (this holds the Chemcassette cartridge in place)� Take the detector out of
Monitor mode before opening the door�
NOTE
Chemcassette cartridges must be stored according to the manufacturer's
guidelines when not in use�
Chemcassette
cartridge
Encoder
Optics gate card
Figure 7. The optics gate card and Chemcassette cartridge
1� If the detector's handle is installed, release it by unscrewing the thumbscrew�
For fixed installations, skip to Step 3�
2� Tilt the handle up as far as possible�
3� Unfasten the latches on the sides of the detector (two on each side) that
secure the Chemcassette cartridge access cover�
4� The detector has slotted hinges� Lift up the cover slightly and then open it all
the way to the right�
5� Select the Change Chemcassette menu option and follow the on screen
instructions�
6� If required, remove the old Chemcassette cartridge� When prompted, snap
the new Chemcassette cartridge into place. It will fit in only one orientation.
The takeup spindle will "flutter" back and forth to allow easy installation.
7� To close the cover, position it over the base and press so that it contacts the
gasket uniformly�
8� Close the latches on the sides of the detector to secure the cover�
SPM Flex Gas Detector
Take-up spindle
17