Print Options

Results for the Drugs of Abuse (DOA) assays can be printed
using one of the following options:
Factor Mode—Numeric Results
Factor Mode—POS/NEG Results
Absorbance Mode
Factor Mode—Numeric Results
When the system is selected for Quantitative Ranges on the
Outline page of the Assay Configuration screen, results print
a positive number for a value equal to or higher than the Multi
DOA Cutoff calibrator, and a negative number for a value
lower than the Multi DOA Cutoff calibrator. A positive number
is interpreted as positive; the specimen contains drug. A
negative number is interpreted as negative; the specimen does
not contain drug, or drug present yields a response less than
that of the cutoff concentration for the assay being tested.
NOTE: Positive and negative numeric response values
do not represent drug concentrations.
Perform the following steps to configure an assay in Factor
mode—Numeric.
1. Select <Assays> in the Information Access Area of the
Main Display. The ASSAYS screen displays.
2. Select the desired assay button. The ASSAY STATUS
screen displays.
3. Select <Config> in the right column of the screen. The
Assay Configuration screen displays.
Assays/Reagents
23