AEROSET
System Troubleshooting Guide
94816-107—November 2004
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution

Procedure

1. Prepare one blank level for each assay to be tested.
2. Select <Database> in the Information Access Area of the
Main Display. The DATABASE screen displays.
3. Select <Order> in the right column of the screen. The
Order Samples screen displays. If this is not the first
sample ordered, select <New>.
4. Enter the following information:
Name or SID
C/P
Replicates (5)
5. Select the assay(s) to be evaluated.
6. Select <Copy> in the right column of the Order Samples
screen. The Copy Order dialog window displays.
7. Enter 1 in the How Many Copies? field to request one copy
(N = 10, 2 samples × 5 replicates) then select <OK>. The
Order Samples screen displays.
8. Load the sample(s) in the designated C/P.
9. Select <OK> in the right column of the screen.
10. Select <RUN> in the Action Area of the Main Display. The
RUN OPTIONS screen displays.
11. Select the Patient option ( ).
12. Select <Start> in the right column of the RUN OPTIONS
screen to initiate the run.
Assays/Reagents
93