Error Messages
Troubleshooting
Disposal
Mains fuse
To protect the device from supply problems, the Cryo 6 is equipped with a 2-pole
triggers
overload fuse integrated into the main switch (10). If the fuse is triggered, the device
automatically switches off via the toggle switch (10). To restart the device, it must be
switched back on using the toggle switch (10).
If the fault occurs frequently and successively, please notify customer service.
Decreased cooling
If the cooling capacity decreases and the air current is significantly reduced, a
capacity / reduced
polluted defrost drain (8) may be the cause. Dust particles collect in the heat
cooling capacity
exchanger through the treatment air. After a long time, these can clog the defrost
drain and can lead to the condensate backing up.
The defrost drain is above the container for defrosted water. To clean the defrost
drain; the device has to be defrosted (see chapter 7 "Defrosting"). After defrosting,
switch the device off and unplug the power supply from the outlet.
Remove the container for defrosted water and place a flat container beneath the
device to collect the condensate. Slowly unscrew the defrost drain.
Clean the defrost drain under running water and screw it back in.
Error Messages
When switching on the device and sometimes during ongoing operation, device
components important for the program sequence are tested. If an error is
determined, the treatment will be canceled, a fault notification appears in the top line
of the display and an acoustic signal will sound.
The error message appears in clear text in the top line of the display. If treatment is
no longer possible, the device can only be switched off.
Faults that do not impact further operation of the device can be repaired by switching
the device off, waiting five seconds and switching it back on.
To remove fault notifications for overheating or excess pressure, the device should
only be switched on after 30 minutes because the device requires time to cool down.
The external and ambient temperatures, which impact the cooling capacity, can be
the cause for this.
If the fault occurs frequently and successively, please notify customer service. You
can reach customer service through your assigned field technician or the central
office in Neu-Ulm.