Safety and Maintenance

14.2 Maintenance

Cleaning the air filter
In order to clean the treatment air and necessary cooling air from dust particles, a
filter element (9) is located on the rear of the device. Under normal use it is sufficient
to clean the filter regularly by vacuuming the entire surface of the outer area using a
commercial household vacuum cleaner, at the latest, when the software displays a
corresponding request notification after 200 operating hours. In carpeted rooms or
rooms in which other circumstances lead to increased dust generation, more
frequently cleaning the filter element is recommended.
Notice:
After cleaning the air filter, the notification from the software must be confirmed by
pressing the "OK" key.
Emptying the container
If the Cryo 6 is switched off after use or the defrost program is started, the cooling
for defrosted water
system heats up and generates condensate.
The one liter container for defrosted water can be removed for draining by pulling it
out of the device.
The container must be put back into the device.
Please also comply with the specifications for cleaning and disinfecting (Chapter 10).