Profibus Address - Dräger Polytron 8100 Instructions For Use Manual

Hide thumbs Also See for Polytron 8100:
Table of Contents

Advertisement

Menu
Set current
This function sets the current to any value between 0 and
22 mA.
Select Settings > Communication > Analog interface >
Set current and confirm.
Once the alarms at the central controller are inhibited,
confirm the message Inhibit all alarms.
Select the line for editing the current and confirm.
Set the current and confirm.
The setting for the current is displayed.
Select Next and confirm.
Select Set current out or Current off and confirm.
Select Next and confirm.
The function is terminated.
Once the alarms at the central controller are enabled
again, confirm the message Enable all alarms.
Set concentration
This function sets the current corresponding to any
concentration value between 0 and full-scale.
Select Settings > Communication > Analog interface >
Set concentr. and confirm.
Once the alarms at the central controller are inhibited,
confirm the message Inhibit all alarms.
Select the line for editing the concentration and confirm.
Set the concentration and confirm.
The setting for the concentration is displayed.
Select Next and confirm.
Select Set concentr. out or Concentration off and
confirm.
Select Next and confirm.
The function is terminated.
Once the alarms at the central controller are enabled
again, confirm the message Enable all alarms.
Set fault
This function sets the current to the fault current.
Select Settings > Communication > Analog interface >
Set fault and confirm.
Select Enable or Disable and confirm.
Set warning
This function sets the current to the warning current.
Select Settings > Communication > Analog interface >
Set warning and confirm.
Select Enable or Disable and confirm.
Set maintenance
This function sets the current to the maintenance current.
Select Settings > Communication > Analog interface >
Set mainten. and confirm.
Select Enable or Disable and confirm.
24
Set beam block signal
This function is only available for Polytron 87X0 (using the
sensor PIR 7X00).
5.8.3

Profibus address

Not yet active - for future functions.
5.9
Settings sensor
5.9.1
Sensor change
With this function, a sensor can be changed during operation
without triggering a fault signal at the central controller. It also
ensures that all sensor data in the microprocessor can be
saved in the sensor memory (EEPROM) before disconnecting
the sensor.
A sensor can be changed at any time. However, for safety
reasons a fault alarm is generated when the sensor is
disconnected until a new sensor is connected.
1. Select Settings > Sensor > Sensor change and confirm.
The maintenance signal is generated on the analog
interface, the following display shows: "Please remove the
sensor".
2. Replace the old sensor with a new sensor, see Section 6.4
on Page 33.
When the new sensor is installed:
Display: "Loading data, please wait".
When the sensor data are loaded:
Display: "Data loaded".
3. Select Back to menu and confirm.
The maintenance signal on the analog interface remains
pending until the sensor is warmed-up.
The warm-up time depends on the type of sensor and its
history. For example, if the sensor has already warmed-up on
another instrument and the operation was only interrupted for
a short time, the warm-up time can be shorter than the time
specified in the Instructions for Use for the sensor.
Maximum warm-up time of a new sensor: see the Instructions
for Use for the sensor.
NOTICE
i
i
If a sensor of the same type (same part number) was
previously
installed,
configuration is retained (gas type, measurement
range, test gas, calibration interval, etc.)
Otherwise the factory default settings of the new
sensor (see Instructions for Use for the sensor) are
uploaded and will overwrite the instrument specific
configuration.
This can be prohibited if the sensor lock function (see
Section 5.9.4 on Page 25) is activated.
the
instrument
specific
Dräger Polytron 8100

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents