Figure 10.3. Hours of Use screen.
Table 10.1.
Daily Maintenance
As required, clean all exterior surfaces, excluding the display with a light application of
disinfectant and wipe dry with paper towels.
NOTE: DO NOT use solutions containing ammonia.
Hardware computer components should be wiped clean as needed using a soft rag dampened
with alcohol.
Quarterly Maintenance
Lubricate the treadbelt and slider deck. The Biodex Lubricant Kit is designed to reduce friction
between the treadbelt and the slider deck. It is required for all institutional treadmills. Proper
90
Hours of Usage Guidelines
Instruction
Lubricate Deck
Reverse Deck
Belt Replacement
Clean Motor and Amp
Front Roller Cleaning
Hours of Usage
75
1,000
1,000
750
375