Meaning Of Symbols In The User Interface; Displaying The Last Measured Values; Measurement Series And Storage - Gossen MetraWatt SECUTEST PRO Manual

Test instruments for measuring the electrical safety of devices
Hide thumbs Also See for SECUTEST PRO:
Table of Contents

Advertisement

8.2

Meaning of Symbols in the User Interface

Sym-
Softkey Variants, Single Measurements
bol
Set parameters
Accept changed parameters, acknowledge memory location
Acknowledge messages during tests/measurements
or resume test sequence
Abort measurement
Direct selection key for selecting the measurement type
Currently selected polarity: "L-N"
Press key to change polarity
Currently selected polarity: "N-L"
Press key to change polarity
Ip
Direct selection key for selecting test current for protective
conductor measurement
U+
Direct selection key for changing voltage in 10 V steps for
U–
insulation measurement
Start evaluation – record measured value. Each time this
softkey is pressed, an additional measured value is saved
and the number is increased by one.
The ID number to which the measurement(s) will be stored
can be entered here.
Valid measured values have been obtained for a measure-
ment. This measurement can be saved.
Save measurement data as (with display of directory path
/ ID or new entry of an ID other than the preselected one)
Display measured values from performed measurements
Magnifying glass symbol: show (+) or hide (–) details regard-
ing database objects or selected measurements
8.3

Displaying the Last Measured values

1 Start the measurement by pressing the
The symbol shown at the right appears and indicates
how many measurements have already been per-
formed.
2 Stop the measurement by pressing the
key, unless a specified measuring time has been stipu-
lated.
The save symbol (floppy disk with a number 1)
appears and indicates that one valid measured value has
been recorded, which can now be saved.
3 Press the save icon (floppy disk).
"No DUT selected!" appears.
4 In order to view the last measured values, press the
symbol for executed measurements after testing. The
last measured values are displayed.
5 The desired measurement can be selected with the
scroll keys.
6 The associated measuring parameters can be shown or
hidden using the keys shown at the right.
7 The measured value view is exited by pressing the
green checkmark, in order to subsequently save the mea-
sured values (as described in Section 8.4) or to return to the
initial view by pressing the ESC key.
8.4

Measurement Series and Storage

Single measurements can be combined into measurement series.
The measured values can be saved by pressing the save key, or
measurement series can be generated. These can be saved to a
DUT (ID number) which has already been set up in the database
(see Section 5.4.1). The appearance of the save key changes
depending on meaning.
22
Measuring Sequence with Pre-Selection of the DUT
1 Activate the database view (MEM navigation) by pressing the
2 Select the DUT or its ID number for the following measure-
3 Return to the measuring view by pressing the MEM key or
4 Start the test with the
5 Each time the key at the right is pressed, the respec-
6 Stop the measurement by pressing the
7 If you press the save symbol now (floppy disk), the dis-
8 After pressing the save symbol once again, acknowledgement
Measuring Sequence with Subsequent Entry of the DUT
1 Start the measurement by pressing the
2 Stop the measurement by pressing the
3 Press the save symbol (floppy disk).
4 You are informed that you haven't selected a DUT in
START/STOP
key.
5 There are two ways to subsequently select a DUT
START/STOP
1
6 The cursor jumps to the location of the DUT with the selected
1
7 Press the save symbol (floppy disk).
MEM key.
ments with the scroll keys.
the
START/STOP
key.
START/STOP
The symbol shown at the right appears and the zero
indicates that no measurements have yet been
recorded or saved to buffer memory.
tively current measured value is saved to buffer mem-
ory and the number shown in the symbol is increased.
In this way, you always know how many measurements have
already been recorded.
key, unless a specified measuring time has been stipu-
lated.
The save as symbol appears (floppy disk icon with the
number of measured values saved to the clipboard).
play is switched to the DUT in the database view for checking.
of successful storage appears. At the same time, the display
is switched to the measuring view.
The symbol shown at the right appears and indicates
how many measurements have already been per-
formed.
key, unless a specified measuring time has been stip-
ulated.
The save symbol (floppy disk with a number 1)
appears and indicates that one valid measured value has
been recorded, which can now be saved.
the database.
using an ID number which has already been set up in
the database:
– Select the ID number with a barcode scanner
or
– Enter an ID number by pressing the ID key.
ID number. You only need to acknowledge this position by
pressing the green checkmark.
A message appears indicating that the data have been suc-
cessfully saved and the display is switched to the measuring
view.
Note
If the entered number cannot be found in the database
(because it hasn't been set up), it can be entered imme-
diately by pressing Yes when the prompt appears.
However, the storage location cannot be selected in this
case. The measurement is saved to the most recently
selected hierarchy.
Note
Measurements and measurement series can only be
saved after measurement has been completed. Mea-
sured values can only be added to intermediate buffer
memory during a measurement. Customer, location and
other entries cannot be changed in the memory menu.
These have to be selected directly in the database and
entered or changed.
key.
0
1
START/STOP
3
START/STOP
key.
START/STOP
3
3
GMC-I Messtechnik GmbH

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SECUTEST PRO and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Secutest base 10

Table of Contents