AND HV-15KC Instruction Manual page 37

Digital platform scale
Table of Contents

Advertisement

Calibration test report
The calibration test mode is used to compare a calibration weight with the calibration
test data weighed by the scale.
 This test does not perform calibration.
Step 1. In the weighing mode, press and hold the
CAL switch until CC
the switch.
The calibration test mode is not available when
the function setting "inf1 0" or "inf2 0" is
selected.
Step 2. Press the ENTER key to display CC 0
Step 3. If necessary, change the value of the calibration
weight as described in "13.2.4. Span Calibration"
Step 4. With nothing on the pan, press the
key. The zero point is measured and the
measured value with the unit "kg" is displayed for
a few seconds. Then, the value of the calibration
weight is displayed.
Step 5. Place a weight of the same value as displayed
on the pan and press the
measure it. The measured value with the unit
"kg" is displayed for a few seconds.
6.
end
Step
gmp
7.
Step
output.
8.
CC
Step
Press the ON/OFF key to turn the power OFF
or press the CAL switch.
HV/HW-C/CP Series
appears, and release
appears.
appears and calibration test report is
appears again. Remove the weight.
.
ENTER
ENTER
key to
Page 35
CAL
Release the keys
Change the calibration
weight value if necessary
Weighing unit
Weighing unit
To output the results