VPAP ST and VPAP S technical specifications

Pressure and
flow state
Mode reading
Sound:
DECLARED
DUAL-NUMBER
NOISE
EMISSION
VALUES in
accordance with
ISO 4871:1996
Physical
Air filter
Supplemental
oxygen
H5i technical specifications
Temperature
Physical
Material
32
Operating pressure range (measured at the mask): 4–20 cm H
2–25 cm H
O (VPAP)
2
Maximum single fault steady state pressure: 30 cm H
for > 6 sec; 40 cm H
2
CPAP mode
Set pressure: 4 to 20 cm H
ST and T modes
IPAP: 4 to 25 cm H
O; EPAP: 2 to 25 cm H
2
Sound pressure level (CPAP mode)
with SlimLine tube: 26 dBA with uncertainty of 2 dBA as measured
with Standard tube: 27 dBA with uncertainty of 2 dBA as measured
with either SlimLine tube or
Standard tube and H5i:
Sound power level (CPAP mode)
with SlimLine tube: 34 dBA with uncertainty of 2 dBA as measured
with Standard tube: 35 dBA with uncertainty of 2 dBA as measured
with either SlimLine tube or
Standard tube and H5i:
Dimensions (L x W x H): 153 mm x 140 mm x 86 mm
Weight: 835 g
Housing construction: Flame retardant engineering thermoplastic
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Standard: Polyester non-woven fiber
Hypoallergenic: Acrylic and polypropylene fibers in a polypropylene carrier
Recommended maximum supplemental oxygen: 15 L/min
Maximum heater plate temperature: 65°C
Temperature cut-out: 74°C
Maximum gas temperature at mask: ≤ 41°C
Dimensions (L x W x H): 153 mm x 145 mm x 86 mm
Weight (standard water tub): Docking station and unfilled water tub 690 g
Weight (cleanable water tub): Docking station and unfilled water tub 790 g
Water capacity: To maximum fill line 380 mL
Docking station: Flame retardant engineering thermoplastic, aluminium
Cleanable water tub: Injection molded plastic, stainless steel and silicone seal
Standard water tub: Injection molded plastic, aluminium and elastomer seal
O - if pressure exceeded for >1 sec
O
2
2
according to EN ISO 17510-1:2009
according to EN ISO 17510-1:2009
28 dBA with uncertainty of 2 dBA as measured
according to EN ISO 17510-1:2009
according to EN ISO 17510-1:2009
according to EN ISO 17510-1:2009
36 dBA with uncertainty of 2 dBA as measured
according to EN ISO 17510-1:2009
O - if pressure exceeded
2
O
O (CPAP);
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

H5iVpap s

Table of Contents