Technical Specifications; General Technical Specifications - ResMed VPAP ST Clinical Manual

Positive airway pressure device, heated humidifier
Hide thumbs Also See for VPAP ST:
Table of Contents

Advertisement

Technical specifications

General technical specifications

Power supply
90W power supply unit
Input range: 100–240V; 50–60Hz; 115V, 400Hz nominal for aircraft use
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
30W power supply unit
Input range: 100–240V; 50–60Hz; 115V, 400Hz nominal for aircraft use
Typical power consumption: 20W (40VA)
Maximum power consumption: 36W (75VA)
90W DC/DC converter
Nominal inputs: 12V, 24V
Typical power consumption: 70W
Maximum power consumption: 110W
Environmental
Operating temperature: +5ºC to +35ºC
conditions
Note: The air flow for breathing produced by this therapy device can be higher
than the temperature of the room. Under extreme ambient temperature conditions
(40ºC) the device remains safe.
Operating humidity:10–95% non-condensing
Operating altitude: Sea level to 2,591 m
Storage and transport temperature: -20ºC to +60ºC
Storage and transport humidity: 10–95% non-condensing
Electromagnetic
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2:2007 , for residential, commercial, and light
compatibility
industry environments.
It is recommended that mobile communication devices are kept at least 1 m away
from the device.
Information regarding the electromagnetic emissions and immunity of these
ResMed devices can be found on www.resmed.com, on the Products page under
Service and Support. Click on the PDF file for your language.
Aircraft use
ResMed confirms that these VPAP devices meet the Federal Aviation
Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all
phases of air travel.
IEC 60601-1
Class II (double insulation), Type BF
classification
Pressure sensor: Internally located at device outlet, analog gauge pressure type,
Measuring and
display devices
-5 to +45 cm H
Flow sensor: Internally located at device inlet, digital mass flow type,
-70 to +200 L/min
O
2
31
English