Voltage Measurement Accuracy - HP 54501A Service Manual

100 mhz digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

Table 3-2. HP 54501A Performance Test Record
HP 54501A PERFORMANCE TESTS
Hewlett-Packard
Tested
by
Model
54501A
100 MHz Digitizing Oscilloscope
Work Order No .
Date Tested
Serial No .
Recommended Calibration
Interval
12
Months
Para
Test
Test
Limit
Results
No
CHI
CH2
CH3
CH4
3-5
BANDWIDTH
<3 dB @ 100 MHz
>5 .6 div of deflection
-
-
-
-
3-6
OFFSET
< 1 .8 div
-
ACCURACY
< 0 .4 div (5mV range)
-
3-7
TRIGGER
5 mV @10 MHz
0 .8 div
-
-
SENSITIVITY
5 mV @50 MHz
2 div
-
200 mV @10 MHz
0 .4 div
-
200 mV @50 MHz
1 div
-
3-8
DELTA-T
40 nsec ± 1 .3 nsec
ACCURACY
(38 .7 to 41 .3 nsec)
3-9
VOLTAGE
V/div
Offset
Minimum
CHI CH4
Maximum
MEASUREMENT
ACCURACY
5 V
+15 V
27 .94 V
-
-
32 .06 V
2 V
+6 V
11 .18 V
-
-
12 .82 V
1 V
+3 V
5 .58 V
-
-
6 .41 V
500 mV
+1 .5 V
2 .79 V
-
-
3 .21 V
200 mV
+600 mV ,
1 .12 V
-
-
1 .28 V
100 mV
+300 mV
559 mV
-
-
641 mV
50 mV
+150 mV
279 mV
-
-
321 mV
20 mV
+60 mV
111 .8 mV -
-
128 mV
10 mV
+30 mV
55 .9 mV
-
-
64 mV
5 mV
+15 mV
27 .8 mV
-
-
32 .2 mV
CH2 CH3
500 mV
+1 .5 V
2 .79 V
-
-
3 .21 V
200 mV
+600 mV
1 .12 V
-
-
1 .28 V
100 mV
+150 mV
559 mV
-
-
641 mV