HP 54501A Service Manual page 15

100 mhz digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

Table 1-2 . HP 54501A Operating Characteristics
HP 54501A GENERAL INFORMATION
SETUP AIDS (continued)
measurements, may be remotely programmed via
HP-IB (IEEE 488) .
HP-113 programming complies
Save/Recall : Four front-panel setups (1-4) may be
with IEEE 488 .2-1988 "Standard Codes, Formats,
saved in non-volatile memory .
Protocols, and Common Commands" .
Recall Clear . Resets the HP 54501A to Its factory
Data Acquisition and Transfer Rate : A 500-point
default settings .
data record can be acquired and transferred to a
computer at a rate of approximately 10 times per
Recall 0:If Autoscale, ECL, or TTL preset, or
second, as tested with an HP 9000, Series 200
RECALL SETUP are Inadvertently selected,
Controller . The acquired data was type normal,
RECALL 0 restores the memory to Its last state
completion criteria 100% with the oscilloscope at a
prior to selection .
time/div setting of 5
A
s/dIv .
Show : Displays Instrument status, Including
Data Transfer Rates: Approximately 120 Kbytes/s .
volts/div, offset, and trigger condition .
Probe Compensation Output : An approximately
HARDCOPY : The CRT display, Including menus
1 .5 MHz signal Is provided for probe compensa-
and measurement results, can be transferred
tion .
A probe-to-BNC adapter is used to connect
directly to an HP-16 graphics printer, including the
the probe to the rear panel Probe Compensation
HP 2225A ThInkJet® or compatible printers .
BNC output . During calibration, this output is used
to provide other calibration signals.
HP-IB PROGRAMMABILTITY
DC Calibrator Output : This output is used for ver
Full Programmability: The HP 54501A provides
tical calibration of the HP 54501A .
full programmabilty . Instrument settings and
operating modes, including automatic