HP 54501A ADJUSTMENTS
4-5. POWER SUPPLY ASSEMBLY ADJUSTMENT
Required Equipment
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 6114A
Digital M ultimeter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3478A
1 .
Disconnect power cord from HP 54501 A. Refer to figure 4-1 for testpoint and adjustment
locations .
2.
Connect the common lead of the voltmeter to Power Supply Assembly chassis.
3.
Connect the positive lead of the voltmeter to +5.15V on the Power Supply Assembly .
4.
Connect the power cord to the HP 54501A and put power switch in ON position .
5 .
Voltmeter should indicate voltage within the range of +5.150 V to +5 .175 V and not to exceed
upper limit of 5.2 V.
6 .
If voltmeter reading is out of this range, adjust +5 ADJ on Power Supply Assembly to +5 .150 V
± .050 V (+5.125 V to +5 .175 V).
COM- +12V +12V -12V -5 .2V +3 .5V +5 .15V +5.15V ADJ
000000000,
" ! """,0,001
Figure 4-1 . HP 54501A Power Supply Assembly Testpoints and
Adjustment Location