HP 54501A Service Manual page 30

100 mhz digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

HP 54501A PERFORMANCE TESTS
5 .
Repeat steps 1 through 4 for channels 2 through 4.
6 .
Adjust Power Supply for +50 mV output.
7 .
Press HP 54501A front-panel CHAN key.
a.
Select channel 1 on .
b.
Select 5 mV/div .
c .
Set coupling to dc .
d .
Set offset to +50 mV .
8.
Press HP 54501A front-panel display key .
a.
Select avg mode .
b .
Set #of avg to 2048 .
9 .
Check for baseline within 0.6 divisions of center reference .
10 .
Repeat steps 7 through 9 for channel 4.
NOTE
This performance test is temperature sensitive. Test limits in steps 4, 8, and
9 are with instrument operating at same temperature as when calibrated. If
temperatures are not the same, the error will be increased by 0.075 divIA °C
(0.15 div/0° C on 5 mV/div range) .