HP 54501A Service Manual page 72

100 mhz digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

HP 54501A SERVICE
and the HP-113 in accordance with IEEE 488
standards . The 75160 is an 8-bit data buffer
and the 75162 is an 8-bit control line buffer .
The HP-IB is a 24-pin shielded cable carrying
8 data lines, 8 control lines, 7 system grounds,
and 1 chassis ground.
The keypad/RPG interface is interpreted as an
8 X 8 matrix with 8 row lines connected
through an open-collector buffer to the 8 lower
address lines and the 8 column lines connec-
ted through a tri-state buffer to the 8 lower
data lines .
The acquisition interface consists of 16 func-
tional lines : 3 data latch controls, 4 logic trig-
ger functions, 2 DAC functions, 3 timebase
functions, and four unused selects . Other in-
terface circuitry is made up of buffers for ad-
dress and data lines and five latched data lines
for control signals to the Timebase and Logic
Trigger and the calibration outputs to the HP
54501A rear panel . The CPU provides 3
clocks to acquisition via the interface : 1 .5 kHz
to rear panel DC Calibrator Output, 2 .5 MHz
for dither, and 20 MHz to D/A Converter and
Probe Compensation AC Calibrator Output .