Thermocouple Ranges & Accuracies
T/C
D
P
L
J-DIN
U
T-DIN
Quality PIECAL Products Online at:
Based on ≤ ± (0.02 % of Reading + 0.01 mV)
Degrees C
Range
-1.1 to 150.0
150.0 to 1100.0
1100.0 to 1750.0
1750.0 to 2320.0
0.0 to 600.0
600.0 to 900.0
900.0 to 1200.0
1200.0 to 1395.0
-200.0 to -50.0
-50.0 to 300.0
300.0 to 900.0
-200.0 to -50.0
-50.0 to 50.0
50.0 to 550.0
550.0 to 600.0
GlobalTestSupply
www.
°C
Degrees F
Range
±1.0°
30.1 to 302.0
±0.7°
302.0 to 2012.0
±1.0°
2012.0 to 3182.0
±2.0°
3182.0 to 4208.0
±0.3°
32.0 to 1112.0
±0.4°
1112.0 to 1652.0
±0.6°
1652.0 to 2192.0
±0.7°
2192.0 to 2543.0
±0.4°
-328.0 to -58.0
±0.2°
-58.0 to 572.0
±0.3°
572.0 to 1652.0
±0.6°
-328.0 to -58.0
±0.3°
-58.0 to 122.0
±0.2°
122.0 to 1022.0
±0.3°
1022.0 to 1112.0
.com
Page 39
°F
±1.8°
±1.3°
±1.8°
±3.6°
±0.6°
±0.8°
±1.1°
±1.2°
±0.7°
±0.4°
±0.5°
±1.1°
±0.5°
±0.4°
±0.5°
sales@GlobalTestSupply.com