PIE 820-ELITE Operating Instructions Manual page 28

Multifunction process calibrator ma,v,ph,tc,?,rtd,freq,pressure
Table of Contents

Advertisement

RTD Source
Choose this function to provide a simulated RTD signal into
controllers, temperature transmitters, indicators or any input
devices that measure RTD sensors.
Move the power switch
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select RTD for the FUNCTION, °F or °C for the UNITS
and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper
10Ω or 50Ω, Nickel 120Ω or 110Ω curves). Note: Pt 100Ω 3850
is the most common RTD type.
Disconnect all sensor wires from the devices to be calibrated
and connect the PIE 820-ELITE to the inputs of the device using
2, 3 or 4 wires.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob
faster dialing with 10.0° increments.
HI
mA V pH TC Ω RTD Freq Pressure
Quality PIECAL Products Online at:
820.0
α
Pt 100
=3850 °C
SOURCE
MAX
HI
READ
SET
OFF
LO
MIN
READ
READ
Double Click
Menu
Push & Turn
for
Fast Dialing
Push & Hold
to
Store/Step
GlobalTestSupply
www.
to SOURCE then Double Click the
w
e
. Press and turn the knob for
Instrument with RTD Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Page 26
2, 3 & 4 Wire RTD Connections