PIE 820-ELITE Operating Instructions Manual page 35

Multifunction process calibrator ma,v,ph,tc,?,rtd,freq,pressure
Table of Contents

Advertisement

Read mA
Ranges and Resolution
Accuracy
Voltage burden
Overload/Current
limit protection
Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution
Accuracy
Loop compliance voltage ≥ 24 DCV @ 20.00mA
Loop drive capability
mA 2-Wire Transmitter Simulation
Accuracy
Voltage burden
Overload/Current limit
protection
Loop voltage limits
DC Voltage Read
Range and Resolution
Accuracy
Input resistance
Quality PIECAL Products Online at:
Specifications
0.000 to 24.000 mA or -25.00 to 125.00%
of 4-20 mA
≤ ± (0.02 % of Reading + 0.003 mA)
≤ 2V at 24 mA
25 mA nominal
0.000 to 24.000 mA or -25.00 to 125.00%
of 4-20 mA
≤ ± (0.02 % of Reading + 0.003 mA)
1200 Ω at 20 mA for 15 hours nominal;
950 Ω with Hart Resistor enabled
Same as Source/Power & Measure
≤ 2V at 20 mA
24 mA nominal
2 to 60 VDC (fuse-less protected from
reverse polarity connections)
±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00
to 60.00 V DC
≤ ± (0.02 % of Reading + 0.01% Full Scale)
≥ 1 MΩ