PIE 820-ELITE Operating Instructions Manual page 21

Multifunction process calibrator ma,v,ph,tc,?,rtd,freq,pressure
Table of Contents

Advertisement

Read mV/V
Choose this function to measure from -99.999 to 99.999
millivolts, -99.999 to +99.999 mV, 0.000 to 10.250 V dc or 0.00
to 60.00 V dc.
Move the power switch
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select V for the FUNCTION and 10V, 1V, 60V or 100
mV for the RANGE.
Connect the red input lead (+) of the PIE 820-ELITE to the more
positive point of the break and the black input to the more
negative point.
Signals above the maximum scale are limited by protection
circuitry with "OVERRANGE" flashed on the display and the red
OVERLOAD LED lit.
The PIE 820-ELITE measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch
MIN to see the lowest reading. Press and hold the knob
clear the MAX and MIN readings.