ADJUSTMENT & OPERATION
Fig. 13
Fig. 14
®
ACCU-SCOPE
Focusing - (Fig. 13)
Set the left diopter collar ① to "0".
Using your left eye only and the 10X objective, focus your
specimen by adjusting the coarse focusing adjustment
knob ②.
When the image is in view, refine the image to its sharpest
focus by turning the fine focusing adjustment knob ③.
Rotate the diopter collar ① to obtain the sharpest focus.
The diopter range is + or – 5. The number aligned to the
white dot on the viewing head is the diopter in use.
NOTE:
damage to the focusing system
Adjusting the Condenser - (Fig. 14)
The condenser was installed and centered in our factory by
our technicians prior to delivery. If the condenser needs to
be re-installed or adjusted in the future, follow the
instructions on page 13.
To adjust the condenser, turn the condenser focusing
knob ① to raise or lower the condenser ②.
Raise the condenser when using the highest magnification
objective, and lower it when using the low magnification
objective.
Focus the specimen with the 10x objective.
Adjust the condenser focusing knob ① to get a clear
image of the field iris diaphragm.
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
EXC-120 MICROSCOPE SERIES
(continued)
do not counter rotate the focusing knob as this will
.
11