Output Log (Continuous Output); Remote Control Instructions - KERN KFB-TM Operating And Installation Instructions

Force gauge, display devices
Hide thumbs Also See for KFB-TM:
Table of Contents

Advertisement

Symbols:
ST
Stable value
US
Instable value
GS / GW
Gross weight
NT
Net weight
TW
Tare weight
NO
Number weighing processes
TOTAL
Total of all individual weighings
<lf>
Space line
<lf>
Space line
Counting
****************************
PCS
****************************

10.3 Output log (continuous output)

Weighing
HEADER1: ST=STABLE, US=UNSTABLE
HEADER2: NT=NET, GS=GROSS

10.4 Remote control instructions

Command Function
S
Stable weighing value for the weight is
sent via the RS232 interface
W
Weighing value for the weight (stable or
unstable) is sent via the RS232 interface
T
No data are sent, the balance carries out
the tare function.
Z
No data are sent, the zero-display
appears.
P
Quantity will be sent via the RS232-
interface
50
100
Printout examples
ST,GS
1.000KG
US,GS
1.342KG
ST,GS
1.000KG
-
-
10PCS
KFB/KFN-TM-BA_IA-e-1624

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kfn-tm

Table of Contents