8.2.4 P6 output

In this menu item you can select between 5 different types of data output:
Operation
Call up menu item "P6 Output" (see chapt. 7.1)
Press  button
01 Printout no.
Press the arrow keys ( ) to select "01 printout
no."
Press the -key; the current menu item blinks
Press the arrow keys ( ), to select among the
settings below:
0 standard
1 1. printout
4 4. printout
Confirm by pressing the PRINT key
8.2.4.1 Standard data output (0 standard)
Data output is carried out by actuating the PRINT-key.
Definition of the log header is executed in menu item "P2 GLP".
Examples in weighing mode:
Date:
22/10/2013
Time:
13.04.23
User:
Mustermann
Project:
XW/456
0.008 g
40
Display
P1
C
P2 GLP
P3 D
P4 Redout
P5 RS-232
P6
Printouts
P7
Globals
Date:
22/10/2013
Time:
13.16.49
User:
Mustermann
Project:
XW/456
Balance:
10
?
62.685 g *
*: ? = instable measuring value