Adjustable Bracket; Air Purge Collar; Right Angle Mirror - RayTek XR Operating Instructions Manual

Infrared sensor lt,lth,mt,g5,p7
Table of Contents

Advertisement

Accessories

5.2 Adjustable Bracket

Figure 30: Adjustable Bracket (XXXTXXACAB)

5.3 Air Purge Collar

The Air Purge Collar (XXXTXXACAP) is used to keep dust, moisture, airborne particles, and vapors
away from the lens. It can be mounted before or after the bracket. The airflow is passed through 1/8"
NPT stainless steel fitting onto the front aperture. Airflow should be a maximum of 0.5 to 1.5 l/s (0.13 to
0.4 gallons/s). Clean, oil free air is recommended.
Figure 31: Dimension of Air Purge Collar

5.4 Right Angle Mirror

The Right Angle Mirror (XXXTXXACRA) is used to turn the field of view by 90° against the sensor axis.
It is recommended when space limitations or excessive radiation do not allow to directly align the sensor
to the target. The mirror must be installed after the bracket and after the Air Purge Collar and screwed
in fully. In dusty or contaminated environments, air purging is required to keep the mirror surface
clean.
XR
37
Rev. B2 Feb 2017